System Overview
POLISMART Instruments
PRD&RIID PM1801
PRD&RIID PM1802
Hand-Held RIID PM1401K
Decision Support System
POLISMART.NET Service
The POLISMART™ IRCS enables the THE INSPECTOR AND OPERATOR to do the following... More
The POLISMART™ IRCS enables the HEAD of the security department to do the following... More
The POLISMART™ IRCS enables the EXPERT of radiation control to do the following... More
 
Search and Spectrometry gamma-channel PM1401K
Sensitivity, no less
241 Am
137 Cs

200 (cps)/(uSv/h)
200 (cps)/(uSv/h)
Energy range
by a special request
0.06 - 3.0 MeV
0.03 - 3.0 MeV
Setting range of coefficient n (number of standard deviations of background) 1.0 - 9.9
Number of accumulation channels of scintillation spectra 1024
Number of spectra saved in non-volatile memory äî 100
Detection from a distance of 0,2m at moving velocity of 0.5 m/s and background level of no more than 25 uSv/h of gamma radiation sources:

133 Ba
137 Cs
60 Co





55.0 ê Bq
100.0 ê Bq
50.0 ê Bq
Detection from a distance of 0,2m at moving velocity of 0.5 m/s and background level of no more than 25 uSv/h of standard samples:
Pu
U



0.3 g
10 g

Search neutron channel PM1401K
Detector He-3 tube
Energy range From thermal up to 14 MeV
Setting range for coefficient n (number of standard deviations of background) 1.0 - 9.9
Detection at 1 m distance, moving velocity of 0.5m/s and background level of no more than 0.25 uSv/h of the alternative 252 Cf with neutron flux 1,5 õ 104 s-1 which is equivalent to Pu (when using the instrument with the moderator-chamber) 250 g

Measuring gamma channel PM1401K
Detector GM counter
Dose equivalent rate (DER) measurement mode 0.1-105 uSv/h
Energy range 0.015 - 15 MeV
Energy response relative to 0.662 MeV (Cs-137) in the photon radiation measurement mode, no more:

 
-within energy range from 0.015
up to 0.045 MeV
- within energy range from 0.045
up to 15 Ì eV





+40%

+30%
Accuracy of DER measurement (where H is the DER value in uSv/h) +(15+0.0015/H)%

Measuring alpha-beta channel PM1401K
Detector GM counter
Measurement range of α -flux density from 15 cm -2 min -1
up to 10 5 cm -2 min -1
Minimum detectable ? -flux density from 2 cm -2 min -1
Accuracy of α -flux density on 239 Pu ( where γ is the measured α -flux density in ñm-2 min-1) ± (20 + 450/?)%
Measurement range of β -flux density from 6 cm -2 min -1 up to 105 cm -2 min -1
Accuracy of measurement of β -flux density within measurement range on 90Sr+ 90 Y (where γ – the measured β -flux density in cm -2 min -1 ) ± (20 + 60/γ)%

General specifications of the PM1401K
Alarming devices Visual (LCD),built-in audible and external vibration
Communication channels IRDA (IR-channel) and Bluetooth (radio channel)
Battery lifetime 600 h
Power supply AA
Environmental conditions:
- temperature
- relative humidity

-30 ... + 50 °C
up to 95 % at 35°C
Water tightness IP65
Weight 650 g
Dimensions 242 õ 57 õ 55 mm
 
 
 
 